CytoVue SCC Somatic Cell Counter
CytoVue SCC is a benchtop fluorescence imaging somatic cell counter for raw milk. It detects clinical and sub-clinical bovine mastitis and monitors milk quality within minutes — no dilution, microscopy-grade accuracy traceable to ISO 13366 (IDF 148), R² > 0.98 with flow cytometry.
| Detection principle | Fluorescence staining + automatic imaging counting (SCC chip) |
| Sample type | Raw (whole) milk |
| Sample volume | 100 μL |
| Measuring range | 0 – 2,000,000 cells/mL |
| Chamber volume | 4.5 μL per chamber |
| Repeatability | CV ≤ 5% @ 500,000 cells/mL; CV ≤ 7% @ 200,000 cells/mL |
| Correlation | R² > 0.98 with Bentley / FOSS flow cytometry |
| Display | 7-inch touch screen |
| Dimensions | 230 × 215 × 245 mm |
| Weight | 2.7 kg |
| Standards | ISO 13366 (IDF 148); NY/T 800-2004 |
- Clinical & sub-clinical bovine mastitis diagnosis
- Raw milk quality control for dairy plants and collection centers
- Traceable to microscopy gold standard; R² > 0.98 vs flow cytometry
- No dilution required — direct results on a 7" touch screen
- Original & marked fluorescence images stored for every sample
- Compact 2.7 kg benchtop design, easy to move between lab and farm
- OEM / ODM and customized software supported
CytoVue SCC Milk Somatic Cell Counter
The CytoVue SCC is a compact benchtop analyzer that counts somatic cells in raw milk by fluorescence imaging. Used together with single-use counting chips and staining reagent, it delivers laboratory-grade SCC results in minutes — no dilution, no complicated operation.
- Clinical & sub-clinical mastitis diagnosis — detect udder infections early, even when milk looks normal, and manage dairy herd health effectively.
- Raw milk quality monitoring — verify somatic cell counts before collection and processing to safeguard milk quality and food safety.
Fluorescence Imaging Principle
An LED light source passes through a beam-shaping lens and excitation filter to illuminate the sample on the SCC microfluidic chip. Fluorescence signals from stained somatic cells pass through a dichroic mirror, imaging lens and filter set onto a high-sensitivity camera module. The embedded software automatically identifies and counts the cells chamber by chamber — microscopy-grade accuracy without manual work.

Simple 5-Step Workflow — Results in Minutes
- Warm the raw milk sample at 40 °C for 10 minutes.
- Invert and mix the sample thoroughly before pipetting.
- Pipette 100 μL of milk and mix well with the staining reagent.
- Load 8.5 μL into each channel of the counting chip.
- Insert the chip into the analyzer and start the test — the SCC result is displayed automatically.

Simple Software, Complete Data
The 7-inch touch screen keeps daily work effortless: start a test in two taps, register up to 4 samples per run, and review every result with its original and marked fluorescence image. All records stay on the device with built-in search, filters and data export.


Validated Performance
Results are traceable to the reference method — manual microscopy counting with trypan blue staining — following ISO 13366 (IDF 148) and Chinese standard NY/T 800-2004. The instrument was calibrated against Bentley and FOSS flow-cytometry analyzers with more than 500,000 reference samples:
- Correlation with Bentley: R² = 0.9911 (y = 0.9995x + 1.195)
- Correlation with FOSS: R² = 0.9951
- Deviation versus microscopy: < 10%
- Repeatability: CV ≤ 5% at ~500,000 cells/mL; CV ≤ 7% at ~200,000 cells/mL

Typical Applications
Dairy farms · veterinary clinics · milk collection and processing centers · artificial breeding stations · research and teaching laboratories. OEM / ODM service and customized software are available.
Inquire Now
Get price, specifications and shipping cost — reply within 24 hours.