Home / Products / Milk Somatic Cell Counter

CytoVue SCC Somatic Cell Counter

Model: CytoVue SCCMilk Somatic Cell CounterVeterinary

CytoVue SCC is a benchtop fluorescence imaging somatic cell counter for raw milk. It detects clinical and sub-clinical bovine mastitis and monitors milk quality within minutes — no dilution, microscopy-grade accuracy traceable to ISO 13366 (IDF 148), R² > 0.98 with flow cytometry.

Detection principleFluorescence staining + automatic imaging counting (SCC chip)
Sample typeRaw (whole) milk
Sample volume100 μL
Measuring range0 – 2,000,000 cells/mL
Chamber volume4.5 μL per chamber
RepeatabilityCV ≤ 5% @ 500,000 cells/mL; CV ≤ 7% @ 200,000 cells/mL
CorrelationR² > 0.98 with Bentley / FOSS flow cytometry
Display7-inch touch screen
Dimensions230 × 215 × 245 mm
Weight2.7 kg
StandardsISO 13366 (IDF 148); NY/T 800-2004
  • Clinical & sub-clinical bovine mastitis diagnosis
  • Raw milk quality control for dairy plants and collection centers
  • Traceable to microscopy gold standard; R² > 0.98 vs flow cytometry
  • No dilution required — direct results on a 7" touch screen
  • Original & marked fluorescence images stored for every sample
  • Compact 2.7 kg benchtop design, easy to move between lab and farm
  • OEM / ODM and customized software supported
Send InquiryEmail Us

CytoVue SCC Milk Somatic Cell Counter

The CytoVue SCC is a compact benchtop analyzer that counts somatic cells in raw milk by fluorescence imaging. Used together with single-use counting chips and staining reagent, it delivers laboratory-grade SCC results in minutes — no dilution, no complicated operation.

Fluorescence Imaging Principle

An LED light source passes through a beam-shaping lens and excitation filter to illuminate the sample on the SCC microfluidic chip. Fluorescence signals from stained somatic cells pass through a dichroic mirror, imaging lens and filter set onto a high-sensitivity camera module. The embedded software automatically identifies and counts the cells chamber by chamber — microscopy-grade accuracy without manual work.

CytoVue SCC fluorescence imaging detection principle

Simple 5-Step Workflow — Results in Minutes

  1. Warm the raw milk sample at 40 °C for 10 minutes.
  2. Invert and mix the sample thoroughly before pipetting.
  3. Pipette 100 μL of milk and mix well with the staining reagent.
  4. Load 8.5 μL into each channel of the counting chip.
  5. Insert the chip into the analyzer and start the test — the SCC result is displayed automatically.

CytoVue SCC 5-step operating procedure

Simple Software, Complete Data

The 7-inch touch screen keeps daily work effortless: start a test in two taps, register up to 4 samples per run, and review every result with its original and marked fluorescence image. All records stay on the device with built-in search, filters and data export.

CytoVue SCC home interface CytoVue SCC sample reading interface

CytoVue SCC fluorescence image review CytoVue SCC data management interface

Validated Performance

Results are traceable to the reference method — manual microscopy counting with trypan blue staining — following ISO 13366 (IDF 148) and Chinese standard NY/T 800-2004. The instrument was calibrated against Bentley and FOSS flow-cytometry analyzers with more than 500,000 reference samples:

CytoVue SCC versus Bentley flow cytometer correlation

Typical Applications

Dairy farms · veterinary clinics · milk collection and processing centers · artificial breeding stations · research and teaching laboratories. OEM / ODM service and customized software are available.

Inquire Now

Get price, specifications and shipping cost — reply within 24 hours.